GRAS 12AU 1-Channel Universal Power Module with signal conditioning and power amplifier

Connection: Traditional 200 V / CCP
Channel(s): 1

Technology

Typical applications and use

Electroacoustic tests of:

  • Telephones

  • Hearing aids

  • Loudspeakers

  • Portable devices, for example, tablets, GPS, etc.

Design

It is designed for automated production line testing of key loudspeaker parameters, such as frequency response, rub and buzz, THD, and impedance. The 12AU’s design makes it ideally suited for test of portable devices with built-in loudspeaker such as mobile phones, hearing aids, tablets, and GPS devices.

The inherent noise is very low, and by design, it avoids system-generated noise, for example from ground loops. In a measurement setup, it controls the paths of (a) the signal sent out to the loudspeaker and (b) the resulting acoustic signal picked up by a condenser microphone fitted to the preamplifier.

It is remotely controlled via its USB interface and is delivered with a control program for Microsoft Windows ®.

Features

  • Single-channel conditioning amplifier suited for both CCP and 200 V LEMO preamplifiers

  • Very low inherent noise

  • The speaker input and the analog outputs are floating with respect to common ground to avoid system-generated noise

  • Speaker amplifier with measurement of loudspeaker impedance without disturbing the speaker

  • Fulfills EMC requirements according to EU regulations

  • Full remote control via USB interface

  • Gain settings:

    – Microphone conditioner: 0 to 50 dB in 10 dB steps

    – Speaker amplifier: –20 dB to +10 dB in 10 dB steps

    – Current output: 0 dB and +20 dB

  • Filters:

    – Microphone amplifier: 1 Hz or 20 Hz

    – Speaker amplifier: DC or AC coupling (10 Hz)

  • Comprehensive status and overload indication

  • Rack-mountable in 19” rack.

Measurement Control with 12AU

In a complex measurement setup, it is often difficult to avoid system-generated noise. Also, it can be tedious to correctly set up all the parameters of the measurement chain with respect to sensitivity, gain, filtering and overload handling.

With the 12AU, you can control the loudspeaker amplifier and microphone conditioner simultaneously.

The 12AU is remote-controlled via its USB interface and a control program for Windows. You can operate the 12AU using a graphical user interface or by entering commands. Multiple 12AUs can be controlled from the same PC.

These features make the 12AU ideal for automated production line testing.

Technical Specifications

All outputs except the speaker amplifier output have fixed 100 kHz (3 dB) low pass filters to minimize the need for anti-aliasing filtering in the following analyzing equipment. There is no phase shift between microphone input and output due to prepolarized or externally polarized microphones. It is designed without relays.

Warranty

All GRAS products are made of high-quality materials that will ensure life-long stability and robustness. The 12AU is delivered with a 2-year warranty. The warranty does not cover products that are damaged due to negligent use.

Service and Repairs

All repairs are made at GRAS International Support Center located in Denmark. Our Support Center is equipped with the newest test equipment and staffed with dedicated and highly skilled engineers. Upon request, we make cost estimates based on fixed repair categories. If a product covered by warranty is sent for service, it is repaired free of charge, unless the damage is the result of negligent use or other violations of the warranty.

Specifications

CONDITIONING
Frequency range (±3 dB)
Hz
1 to 100 k
Output channel(s)
1
Output impedance
Ω
100
Direct coupling input to output
No
Inherent noise, LIN (20 Hz to 20 kHz), grounded input
µVrms
<1.5 (>= 20 dB Gain)
Inherent noise, LIN with dummy micophone (20 Hz to 20 kHz)
µVrms
< 5
Gain settings, 10 dB steps
dB
0 to + 50
Gain error
dB
< 0.2
High-pass filter
3-pole Butterworth, -1 dB @ 20 Hz
Linear mode
Yes
Type
Constant Current Power (CCP) and Traditional power supply (LEMO)
Input channel, CCP
1
Input channel(s), traditional power supply
1
Preamplifier supply
V
±15
CCP preamplifier supply @ +28 V
mA
4
Polarization voltage
V
0 / 200
Overload indication
Yes
PC controlable
Yes
Monitor output connector
BNC
Monitor output impedance
Ω
100
POWER AMPLIFIER
External filter option
No
Power amplifier input impedance
Ω
10 k
Power amplifer output connector
2-pin LEMO
Power amplifier output current monitor connector
BNC
Power amplifier output current
Ap
1.4
Power amplifier gain settings, 10 dB steps
dB
-20 to 10
Power amplifier frequency range
Hz
DC to 30 k
Power amplifier phase shift response (±1 deg)
Hz
DC to 20 k
Power amplifier output voltage
Vp
±13.5
Power amplifier overload detection
Yes
POWER SUPPLY
Power supply, external (±10 %)
Vac
110 to 240
Battery low indication
No
Fuse
mA
500
Temperature range, operation
°C / °F
-10 to +50 / 14 to 122
Weight
g / oz
2500.00 / 88.18

GRAS Sound & Vibration reserves the right to change specifications and accessories without notice.

Dimensions

12AU dimensions

All dimensions in mm.

Get Quote