G.R.A.S. 42AE Low Frequency Calibrator

Sound pressure level: 140 dB
Freq range: < 0.1 to 250 Hz
ANSI: S1.10
IEC: 60318-1 & -2
  • 42AE Low Frequency Calibrator
  • 42AE Low Frequency Calibrator
  • 42AE Low Frequency Calibrator
  • 42AE Low Frequency Calibrator

Typical applications and use

  • Low-frequency pressure response measurements
  • Low-frequency phase response measurements
  • Comparative measurements
  • Single and dual-channel measurements

Design

The two-port configuration allows the actual sound pressure in the coupler to be monitored by a reference microphone simultaneously with the microphone under test.

The built-in DC-coupled power amplifier enables the calibration to be used for swept-sine, broadband and step function investigations.

The calibrator can be used for both rear-vented and side-vented microphones. In the case of rear-vented microphones, the 42AE can seal the microphone from the preamplifier to obtain the response of the microphone independent of other leaks.

The calibration of microphones in the frequency range below 50 Hz requires a special apparatus which exposes the complete microphone to the pressure variations of the calibration signal, and not just the microphone diaphragm alone. The air equalization system of a microphone is important for determining the low-frequency response, together with the amplifier of the electronic system.

Air equalization must be fast enough to compensate for changes in altitude and barometric conditions, and slow enough not to affect the response of the microphone at the lowest frequencies. Most measurement microphones will be adjusted to a lower limiting frequency of 1 – 3 Hz. But 0.1 Hz is not unusual for infra-sound measurements as well as for airbag testing.

Around the body of the 42AE, which houses the coupler and all its pre-assembled contents, are input and output sockets as well as a green LED.

The accessories included with the 42AE are, more or less, common to all measurements setups. The manual describes a series of configurations and measurement setups and specifies in detail what items are required in each case.

The data sheet shows a setup for exploiting the full potential of the 42AE. It can be used for comparing the low-frequency phase and pressure response of a pair of matched microphones against each other. Alternatively against the known pressure level and phase angle within the coupler of the 42AE.

The voltage signal from the 42AE is analogous to the acoustic pressure and phase angle within the coupler of the 42AE. The relationship between acoustic pressure and voltage output is nominally 1 mV/Pa (the exact value is given on the calibration chart for each individual 42AE).

The dual channel Power Module 12AR is recommended because it can be used at frequencies down to 0.05 Hz.

Frequency range (±2 dB) Hz 0.1 - 250
Input signal, max. Vrms 0.7
ANSI standard S1.10
IEC standard 60318-1 & -2 (former 60318)
Weight g / oz 1.6k / 56.438
Calibration signal 1 mV/Pa (140 dB max)
G.R.A.S. 26AL ¼" Preamplifier
G.R.A.S. 26HG

¼" Preamplifier

G.R.A.S. 40AG ½" Microphone
G.R.A.S. 40AN ½" Microphone
G.R.A.S. 40EN 1" Microphone
G.R.A.S. 12AR Dual-channel Power module